Measurement made easy
Overview
ABB’s Carbon Dioxide Isotope Analyzers meet all of these requirements. In addition, the availability of many value-added options extends the abilities of these units to include discrete samples (collected in bags or vials) and to automatically handle multiple inlet sources.
ABB’s “Enhanced Performance” series incorporates proprietary internal thermal control for ultra-stable measurements with unsurpassed precision, accuracy and drift. Moreover, only ABB’s analyzers provide reliable guaranteed measurements at mole fractions more than 20 times ambient levels.
ABB’s patented OA-ICOS technology, a fourth-generation cavity enhanced absorption technique, has many advantages over conventional Cavity Ringdown Spectroscopy (CRDS) techniques such as being alignment insensitive, having a much shorter measurement time, not requiring tight control of cavity pressure and temperature, and not requiring expensive and power-consuming auxiliary elements.
The analyzer includes an internal computer that can store data practically indefinitely on its internal hard drive (for applications requiring unattended longer term operation), and send real-time data to a data logger through its analog and digital (RS232) outputs. Several optional features are available which improve the flow time response, allow multiple inlet sources, or provide for remote access and control of the analyzer via the Internet.
![]()
Features and benefits
Simultaneous measurements of δ13C:, δ18O:, CO₂ and H₂O
Unprecedented stability, precision and low drift
Measurement rates selectable up to 1 Hz
Installed and operational in minutes
Batch operation via syringe injection option
Insensitive to hydrocarbons or H₂S
Extremely high dynamic range allows measurements of elevated CO₂
Unsurpassed reliability
Real-time diagnostics
Specification (air background)
Precision (1s, 1 second / 10 seconds / 100 seconds):
δ13C: 1.5‰ / 0.5‰ / 0.15‰
δ18O: 8‰ / 2.5‰ / 1‰
12CO2: 0.4 ppm / 0.15 ppm / 0.05 ppm
H2O: 100 ppm (60 sec) / 50 ppm (300 sec)
[Multiply precision specs by two (x2) for Ultraportable]
Maximum Drift (Enhanced Performance models) (peak-to-peak, 1 hr average over 24 hours):
δ13C: < 0.5‰
Total uncertainty (Ultraportable)
< 1%
Measurement Range (meets all specs):
CO2: 380 – 25000 ppm
H2O: 4000 – 60000 ppm
Operational Range:
CO2: 0 – 50000 ppm
H2O: 0 – 70000 ppm (non-condensing)
Measurement Rates:
User-selectable rates up to 1 Hz
Response time to register 95% of a step change:
10 seconds (flow response < 5 sec requires external pump)
Sampling Conditions:
Sample Temperature: -20 – 50 °C
Operating Temperature: 0 – 45 °C
Ambient Humidity: non-condensing (0 – 100% RH)
Fittings:
Inlet: 3/8 in.
Outlet (internal pump): ¼ in.
Outlet (optional external vacuum pump): ½ in.
Outputs:
Digital (RS-232), Ethernet, USB
Power Requirements:
115/230 VAC, 50/60 Hz
66 W [ultraportable model]
350 W [enhanced performance model]
Dimensions (H x W x D):
18cm (7”) x 47 cm (18.5”) x 36 cm (14”) [ultraportable]
28cm (11“) x 97cm (38“) x 56cm (22“) [benchtop]
40cm (15.75“) × 48cm (19“) x 61cm (24“) [rackmount]
Weight:
17 kg [ultraportable]
50 kg [benchtop]
68 kg [rackmount]
Ordering information
LGR-ICOS™ U-CCIA2-915 (Ultraportable)
LGR-ICOS™ CCIA2-912 (Enhanced Performance Benchtop)
LGR-ICOS™ CCIA2-911 (Enhanced Performance Rackmount)
Manufacturer: ABB, Canada